text.skipToContent text.skipToNavigation
  • Cross.svg
    Extensive product ranges
Calling 01685 846666

Spirometer Calibration Log Book

IDZX2015/2

Order from supplier within 5 working days.

This is a non-returnable, special purchase item

£0.01
Add to wish list



The Micro Medical 3 Litre Spirometer Calibration Syringe is designed to simplify the process of spirometer calibration and ensure that devices are functioning correctly.

Used correctly, the syringe holds exactly 3 litres of air, which can be passed through a spirometer in line with its maintenance schedule and guidelines. Each Micro Medical accuracy checking syringe is supplied with a unique certificate confirming ATS/ERS 2005 performance compliance.

Features

• A volume accuracy check should be performed daily
• A 3 litre calibration syringe should be used
• This calibration syringe should have an accuracy of ±0.5% (15 mL for a 3 litre syringe)
• The syringe should be discharged through the device at least three times to give a range of flows varying between 0.5 and 12 litres/second (with 3 litre injection times of approximately 6s for the lower flow limit and less than 0.5s for the upper flow limit)
• The volume at each flow should meet the accuracy requirement of ±3.5%
• Warranty: 2 years

See more See less
Specs
Product Data
Warranty 2 Years
Brand Vyaire
Reviews
Q&A
Downloads